Sampling & Multi Thermal Desorption System
MSTD258/TMD253H

Flow Chart

MSTD258/TMD253H(Multi Thermal Desorption System Manual type)

Features
It is measureement system for out gassing from materials.
Sample (solid sample) is put into the heatr oven of MSTD258M Manual type. press START button swich, then it
goes automatically to heat sample , trap outgas from the sample. After the sampling , remove the sampe tube and
attach it to a thetmal desorption system TMD-253H or another to measure the tube.

Flow Chart
   
MSTD-258M Oven A Type Cat, No.7510-25801 Specitions
- Temp. Range: Ambient -400c
- Quartz chamber: 100X2.8 . or 50mm
- Weight : 55kg approx
- Power Supply : AC100v, 25A
- Gas Hlium : 99095% or higher 400-600 kPa
- Installation Space : 800(w) x 600(D) MM
 
[outline]

- MSTD is an measurement system for out gassing of interior materiaks, construction materials, cleanroom materials, semiconductor materials, foos vessels, or multiple materials which can be a source of indoor air contamination.

- It is to measure a sample put into a chamber in an oven for heating
 

Builing materials out gassing gas screening measurement in indoor air chemicals
(Heating acceleration test)

- Test method of heating materials and outgassing measurement at an acceleerating rate.

- It is used for Comparison or Selecton of outgassing diffusion of builing materials

- It can be used for micro chamber than some L capacity.


Car Interior Materials Sheet ( Heating Temperature 60 )

 
Food Vessels ( Heating Temperature 60 )

 
[ Contaimination from human ]

Dried the 6mm ID Pyrex tube with ultra bath.
Measureed with MSTD-258
Check if blank peaks were not appeared.
Touch the Pyrex tube by hand.
Measure it again with MSTD-258
Fatty acid was detected, it should be come from human skin and squalene were detected very much.
DOP was detected as your concern.

Measure of the Pyrex tube touched by human.

Instrument : 1) MSTD-258
2) GC:HP6890
3)MS: JMS-AM || 50

MSTD conditions

Chamber : Quartz 100mID x 30 mmH
Heating temperature : 30c(1min) - 250 c RAte 20 c/min
Trapping time : 30 min Trapping flow volume : 45 ml/min
Purge gas flow volume : 100ml/min
Trap tube : Tenax 100mg

Measurement 4by MSTD-258


[polymer resin]

<< wafer case measurement>>

Cut each of packing sheet, holder, outer box, cassette wafer box to be set into the quartz chamber (100mmID x 30mmH) and measured each item.

This measurement condition was
Trapping time 30min, purge gas flow volume 100ml/min, trapping flow volume 45ml/min


PP(Cassette)

Heating temperature : 60 c


Elastorner (Holder)

Heating temperature : 60 c



Elastomer (packing)

Heating temperature : 60 c

 